Laboratory froth flotation machine automation GUI
It is designed to work with following data acquisition cards:
- ADVANTECH PCI-1713-AE, 16 - channel differential analog input card
- ADVANTECH PCI-1720, 4 - channel analog output card
Following devices are connected (although these can be changed for other models with analog input/output):
- MVXF50A0-2, adjustable frequency AC drive controller (0-10V analog input)
- FMA-A2409, mass flow meter and controler (0-5V analog input and 4-20mA analog output)
- PHP-701, pH meter and controller (4-20mA analog output)
- TXDIN70, pt100 temperature transmitter (4-20mA analog output)
This connection makes it possible to:
- turn flotation test on/off (agitation)
- automatically adjust air flow after conditioning period is over
- measure temperature and pH
- log all of the above mentioned parameters (for each second of test) and save the results as txt or xls file
引用格式
Ivan Brezani (2024). Laboratory froth flotation machine automation GUI (https://www.mathworks.com/matlabcentral/fileexchange/32985-laboratory-froth-flotation-machine-automation-gui), MATLAB Central File Exchange. 检索时间: .
MATLAB 版本兼容性
平台兼容性
Windows macOS Linux标签
Community Treasure Hunt
Find the treasures in MATLAB Central and discover how the community can help you!
Start Hunting!flot_cont/
版本 | 已发布 | 发行说明 | |
---|---|---|---|
1.0.0.0 |