已回答
Data range is changed when using "dicomread" to read singned short PET dicom image
You may need to scale the PET image by the dicom tag RescaleSlope.
Data range is changed when using "dicomread" to read singned short PET dicom image
You may need to scale the PET image by the dicom tag RescaleSlope.
6 years 前 | 0
已回答
Computing SUV PET Dicom image
OsiriX uses the following formula: SUV (g/ml) = Pixel value (Bq/ml) * Weight (kg) / Dose(Bq) * 1000 (g/kg)
Computing SUV PET Dicom image
OsiriX uses the following formula: SUV (g/ml) = Pixel value (Bq/ml) * Weight (kg) / Dose(Bq) * 1000 (g/kg)
6 years 前 | 0
已回答
Dicomread Stripe in Matlab
It is possible that you may need to scale each slice by its dicom value RescaleSlope.
Dicomread Stripe in Matlab
It is possible that you may need to scale each slice by its dicom value RescaleSlope.
6 years 前 | 0